EMSense™ 40

ETS-Lindgren’s EMSenseTM 40 10 MHz to 40 GHz Electric Field Probe is versatile and ideal for various military, commercial and automotive specifications.
ETS-Lindgren's
laser-powered EMSenseTM 40 Electric Field Probe embodies the latest
innovations in isotropic sensor design, low noise and miniaturized electronics.
Designed to be single range reading, the EMSense 40 can read data continuously
over the entire dynamic range. Data values for each axis (X, Y, and Z) can be
read individually or summed. Fiber optic signal and power lines link the RF
field probe to either the EMCenter 2/7 Slot, or as a direct connect to a PC USB
port with the EMCenter 1-Slot.
The EMCenter 2 and 7
Slot Modular RF Platform along with the EMSense 40 interface card can be used
as a Field Monitor in addition to its capability as a system level platform.
The EMCenter 1-Slot
with the EMSense 40 interface card provide laser power and communications for
the EMSense 40 Field Probe. A USB connection to the PC allows for quick and
easy data collection, using ProbeView V software.
- Laser powered for extended testing
- Frequency corrected field values direct from probe
- No need to apply correction factor
- Single probe frequency range - 10 MHz to 40 GHz
- Broad dynamic range – 2 to 1000 V/m
- Sleek stalk design
- Three axis dipole design
- A2LA Accredited Calibration
- Suitable for MIL Standard Specs:
o MIL-STD 461F Radiated Susceptibility (RS) - Suitable for Automotive Specs:
o SAE J1113/27
o GMW 3091/3097/3103
o FORD FMC 1278 - Suitable for Commercial Specs: EN/IEC 61000-4-3 Radiated Immunity
The EMSense 40 probe
utilizes an updated CPU in the EMCenter to support its improved communication
speed. Some EMCenter units carry an older CPU that must be upgraded to support
the EMSense 40 probe and plug-in card. To check compatibility, access the
" Info" menu on the EMCenter main screen and check
slot 8. The compatible processor is 7000-008. If the EMCenter shows processor
7000-007, a processor upgrade is available through the ETS-Lindgren service
center.
Alternatively, a
stand-alone EMCenter is available for EMSense 40 that appears as an additional
EMCenter slot to EMC control software.
Shape
of housing | Stalk
Probe |
Weight | 3.2
oz (91 g) |
Total
electrical measuring volume | 0.4
in (1 cm) |
Electronics
Housing | 1.18
in (3 cm) |
Tube
Diameter | 0.57
in (1.5 cm) |
Sensor
Cover | 0.97
in (2.5 cm) |
Overall
length | 11.8
in (30 cm) |
Environmental Specifications
Temperature range (operating) | 32 °F to 104 °F (0°C to 40 °C) |
Relative humidity (operating) | Relative humidity (operating) |
Optical Specifications
EMSense 10/40 Power and Communication Interface card | Max. 0.5 Watt output at aperture @ 808 nm |
LASER connector | FC/FC |
Data Connector | ST/ST |
Fibers | 200/230 µm HCS, duplex |
Standard fiber length | 10 m optional 20 and 30 m Maximum 100 m |
Safety Specifications
LASER product classification (Power/Communication card) | Class 1M according to EN 60825-1 and EN 60825-2 |
Safety measures | Code (EMCenter) Remote interlock system (EMCenter) LED indications for LASER ON Audible warning signals Redundant closed loop safety system |
Calibration
Factory calibration | Internally stored, ISO17025 calibration |
Accredited calibration | Traceable, accredited calibration |
Dynamic
range | 2
to 1,000 V/m |
Damage
level | 2,000
V/m |
Frequency
range | 10
MHz to 40 GHz |
Frequency
response (with internal correction) | ±
1.2 dB (10 MHz - 40 GHz) |
Resolution | 0.001 V/m 2 - 10 V/m 0.01 V/m 10 - 100 V/m 0.1 V/m 100 – 1000 V/m |
Linearity | ±
0.5 dB ± 0.5 V/m |
Isotropic
deviation | ±
0.5 dB @ 1 GHz |
Number
of antennas | 3
dipoles |
Measurement
speed (X,Y, Z & ETot) | 100
Measurements/s maximum |